RS-232C INTERFACE - Section 6
MTS-1700 SERIES OPERATION AND REFERENCE MANUAL 6-7
CU A002 15A
MANTA TEST SYSTEMS
VFF#
Voltage - Fault Final
•
Sets the fault final voltage
•
For
Φ
-N and 2-
Φ
-N fault modes, specify values in volts. Valid values are 0 -
150.00 (volts)
•
For
Φ
-
Φ
and 3-
Φ
fault modes, specify value in percent. This percentage is the
percentage of the nominal
Φ
-N voltage. Valid values are 0-100.0 (%).
•
In
Φ
-
Φ
fault modes, this percentage setting is only approximate.
•
Note that, for 2
Φ
-N fault modes, this setting changes the
Φ
-N values. However,
the
Φ
-
Φ
value is shown on the MTS-1710 display.
VPO#
Voltage - Postfault
•
Sets the postfault voltage
•
For
Φ
-N and 2-
Φ
-N fault modes, specify values in volts. Valid values are 0 -
150.00 (volts)
•
For
Φ
-
Φ
and 3-
Φ
fault modes, specify value in percent. This percentage is the
percentage of the nominal
Φ
-N voltage. Valid values are 0-100.0 (%).
•
In
Φ
-
Φ
fault modes, this percentage setting is only approximate.
•
Note that postfault must be turned on (using the POF1 command) for this value
to be applied to the outputs in postfault state
•
Note that, for 2
Φ
-N fault modes, this setting changes the
Φ
-N values. However,
the
Φ
-
Φ
value is displayed on the MTS-1710 display.
VDF#
Voltage - Default
•
Sets the default
Φ
-N voltage on power-up
•
Valid values are 0 - 150.00 (volts)
•
This value will take effect the next time the MTS-1710 is turned on, or when
default settings are restored. (see Section 5.8.1 for details).
6.2.5.2 EXAMPLES.
The following examples show command sequences using the voltage programming commands:
Example 1 Simple
Φ
-N Fault
AVA70
<- set prefault amplitudes to 70V
AVB70
AVC70
PRF1
<- prefault on
POF0
<- postfault off
DYN
<- dynamic operation mode
FMD1
<- simulate B-N fault
VFI15
<- fault amplitude 15V
STR
<- initiate fault
Example 2 Simple
Φ
-
Φ
Fault
FMDO
<- simulate A-N fault
AVA70
<- set prefault
Φ
-N amplitudes to 70V
AVB70