MTS-1750 HIGH CURRENT SOURCE - Section 10
MTS-1700 SERIES OPERATION AND REFERENCE MANUAL 10-7
CU A002 15A
MANTA TEST SYSTEMS
FIGURE 10.6 MTS-1710 + MTS-1720 + (3)MTS-1750 CONTROL CONNECTIONS
10.4.2.2 OPERATION WITH THE MTS-1710 + MTS-1720 + (3)MTS-1750s CONFIGURATION
& SINGLE PHASE OUTPUT.
This configuration allows up to 450A single-phase current output.
10.4.2.2.1 Setup
• Make the rear panel connections, as described in Section 10.4.2.1.
• Select I3-PARALLEL current mode on the MTS-1710.
• Make the front panel current output connections, as shown on Figure 10.7. Note that the front panel
current connections and the rear panel control connections must correspond.
• Turn on the MTS-1710, MTS-1720 and MTS-1750s.
A UX OUT PUT S
A UX IN PUT S
M TS-1750 CH A
C OM 3 SL AVE C UR REN T
C OM 2 R S-232C
C OM 1 R S-232C
M AINS
F 15A
SL AVE C UR REN T IN
POWER OFF
BEFORE
CONNECTING/
DISCONNECTING.
ROTATE TO LOCK
BEFORE USE
SIGNA L IN PUT
M
S
Current output cable
MTS-1710 to MTS-1720 Connection Cable
(Manta Test Systems part # 12-2800-00)
120 VA C 6 0HZ
M ADE IN CAN AD A
CONTROL
120 VA C 6 0HZ
M ADE IN CAN AD A
CONTROL
120 VA C 6 0HZ
M ADE IN CAN AD A
CONTROL
MTS-1750
CHANNEL
A
MTS-1750
CHANNEL
B
MTS-1750
CHANNEL
C
MTS-1710/1720 to MTS-1750 Connection Cable
(Manta Test Systems part # 12-3245-00)