SPECIFICATIONS - Section 2
2-2 MTS-1700 SERIES OPERATION AND REFERENCE MANUAL
MANTA TEST SYSTEMS
CU A002 15A
•
Parallel current mode allows paralleling of current outputs of MTS-1710 and MTS-1720 for
up to 90A max, 1000 VA max, 44 Vrms max.
•
I2-Harmonic maximum current of 17A
Output frequency
•
power line (frequency and phase locked)
•
1st through 10th harmonic of power line
•
variable 40.000 - 80.000 Hz (0.001 Hz resolution, 0.01% accuracy)
80.00 - 800.00 Hz (0.01 Hz resolution, 0.02% accuracy)
•
25Hz frequency mode
Four output level settings:
•
off, prefault, fault, postfault
Phase control:
•
Phase between current and any voltage is adjustable from 0 through 360°
•
Adjustment resolution: 0.25°
Auxilary output contacts:
•
Two auxiliary output contacts
•
Configurable as NO, NC, 52A, 52B (Aux. contacts 1 & 2), Permissive or Unblocking with
programmable delay (Aux. contact 1 only)
2.3 METERING
All outputs are directly metered. This allows direct readout of
Φ
-
Φ
and
Φ
-N voltages, currents and
interphase angles.
Time, frequency, phase, current and voltage are measured simultaneously. In addition, the phase is always
measured between the monitor voltage and the current, allowing all standard phase angle relations to be
directly displayed.
•
AC Voltage measurements: A-N, B-N, C-N, A-B, B-C, C-A
True RMS responding, autoranging
Voltage scales (approx):0-82V, 82-327V
Accuracy: ±0.5% of reading ±0.2% of scale
•
AC Current measurements:
Measures actual output current
True RMS responding, autoranging
Current scales (approx): 0-8.2A, 8.2-32.7A
High current scales (approx): 0-24.5A, 24.5-98.3A
Accuracy: ±0.5% of reading ±0.2% of scale
• DC Current measurement:
Measures actual output current:0 - 10.00A scale
Accuracy: ±1% of reading ±0.2% of scale
•
Phase measurement:
Accuracy: ±0.5° for input levels >5% of range
0 -359.9° or 0 - ±180° display modes
Selectable V-leads-I or I-leads-V measurement reference