MTS-1720 TWO CHANNEL CURRENT SOURCE - Section 9
MTS-1700 SERIES OPERATION AND REFERENCE MANUAL 9-5
CU A002 15A
MANTA TEST SYSTEMS
9.3.3 Getting Started
9.3.3.1 CONNECTIONS AND POWER.
•
Turn off the MTS-1710 and MTS-1720. Connect the main power and safety ground to both the MTS-
1710 and MTS-1720.
•
Connect the [COM3 SLAVE CURRRENT] port of the MTS-1710 to the MTS-1720 [COM1 TO MTS-
1710 PORT], using the provided cable. (Use only a Manta Test Systems supplied cable. This is NOT a
standard RS-232 cable).
•
Connect the MTS-1720 current output to the MTS-1710 slave current input using the high current cable
provided. (On older systems, the connector has a blue locking ring. Retract the blue locking ring before
inserting. After inserting the connector in the mating socket, twist clockwise to lock. Turn the blue inner
locking ring clockwise to lock the connector in place).
•
Turn on the MTS-1710 and MTS-1720 (in any order).
FIGURE 9.3 REAR PANEL MTS-1710/MTS-1720 INTERCONNECTIONS
Power cable portion
MTS-1710 to MTS-1720 Connection Cable Assembly
(Manta Test Systems part # 12-2800-00)
Control cable portion
Connector for
multi-system sync
capability
Connector for
multi-system sync
capability
AUX OUTPUTS
AUX INPUTS
COM 3 SLAVE CURRENT
COM 2 RS-232C
COM 1 RS-232C
MAINS
F15A
SLAVE CURRENT IN
POWER OFF
BEFORE
CONNECTING/
DISCONNECTING.
ROTATE TO LOCK
BEFORE USE
SIGNAL INPUT
MTS-1750
S
M