RS-232C INTERFACE - Section 6
MTS-1700 SERIES OPERATION AND REFERENCE MANUAL 6-11
CU A002 15A
MANTA TEST SYSTEMS
IRR#
Current - Ramp Rate
•
Sets the current ramp rate.
•
Valid values are 0 - 350.0 (amps/sec).
•
A value of 0 turns off current ramping.
•
In I3-PARALLEL current mode, this command sets the current ramp rate for the
paralleled three channels (MTS-1710+MTS-1720 and optional three MTS-
1750s) output.
−
All three current channels must be paralleled together to obtain the
programmed rate of ramp.
−
Valid values are 0 - 350.0 A/sec.
IDU# Current
Duration
•
Set the current fault duration.
•
Valid values are 0 - 99.999 (seconds).
•
A value of 0 turns off current fault duration.
AI2#
Amplitude I2
•
Set the amplitude of I2 in I1&I2 or I2-HARMONIC current modes.
•
Valid values are 0-30 in I1&I2 and 0-17 in I2-HARMONIC.
•
In I2-HARMONIC current mode, this sets the amplitude of the fundamental
component of the output current.
%HM#
Percent Harmonic
•
Set the percentage harmonic of the I2 current output in I2-HARMONIC current
mode in the FAULT state.
•
Valid values are 0 - 50.0 (%)
•
Settable only when AHW# and AHP# are set to zero.
%HP#
Percent Harmonic Prefault
•
Set the percentage harmonic of the I2 current output in I2-HARMONIC current
mode in the PREFAULT and POSTFAULT states.
•
Valid values are 0 - 50.0 (%).
•
Settable only when AHW# and AHP# are set to zero.
AHW#
Amplitude half-wave rectified fault
•
Set the half-wave rectified amplitude of the I2 current output (as measured by an
average responding ammeter) in I2-HARMONIC current mode in the FAULT
state
•
Valid values are 0 - 5.000 Amps.
•
Note: Settable only when %HM# and %HP# are set to zero.
AHP#
Amplitude half-wave rectified prefault
•
Set the half-wave rectified amplitude of the I2 current output (as measured by an
average responding ammeter) in I2-HARMONIC current mode in the
PREFAULT and FAULT states
•
Valid values are 0 - 5.000 Amps.
•
Note: Settable only when %HM# and %HP# are set to zero.