14
TRIGger Subsystem
592
Keysight N8262A Programming Guide
TRIGger[:SEQuence]:LEVel <numeric_value>
This command sets the power level at which a trigger event is recognized.
Syntax
Parameters
Example
Reset Condition
On reset the power level is set to 0 dBm.
:LEV
Space
numeric_value
?
DEF
TRIG
:SEQ
Item
Description/Default
Range of Values
[a]
numeric_value
The power level at which a trigger event is recognized.
–
DEF
: the default value is 0 dBm
Units are resolved to 0.1 dBm.
–40 to 20 dBm
DEF
[a] If a channel offset has been previously set, a higher numeric value is permitted. See
on page 88 for more information.
TRIG:SEQ:LEV 10
This command sets the power level for a
trigger event to 10 dBm.
Summary of Contents for N8262A