SENSe Subsystem
370
Keysight N8262A Programming Guide
Example
Reset Condition
On reset the selected table is not affected.
Query
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:[SELect]?
The name of the selected table is returned as a quoted string. If no table is
selected an empty string is returned.
Query Example
Error Messages
– If
<string>
is not valid, error –224, “Illegal parameter value” occurs.
– If a table called
<string>
does not exist, error –256, “File name not found”
occurs.
– When a sensor calibration table is selected, the power meter verifies that the
number of calibration points defined is one more than the number of frequency
points defined. When a frequency dependent offset table is selected, the
power meter verifies that the number of offset points defined is equal to the
number of frequency points defined. If this is not the case, error –226, “Lists
not the same length” occurs.
– If the
CSET1
command is used when a P-Series or an E-Series Power Sensor is
connected the error –241, “Hardware missing” occurs.
CORR:CSET1 ‘PW1’
This command enters the name of the sensor
calibration table which is to be used on
Channel A.
CORR:CSET1?
This command queries the sensor calibration
table currently used for Channel A.
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...