CALCulate Subsystem
198
Keysight N8262A Programming Guide
CALCulate[1]|2|3|4:LIMit:CLEar:AUTo <boolean>|ONCE
This command controls when the FCO (fail counter) is cleared of any limit failures.
The FCO is used to determine the results returned by the
CALCulate[1]|2|3|4:LIMit:FAIL? query
.
– If
ON
is specified, the FCO is set to 0 each time a measurement is:
– Initiated using
INITiate[:IMMediate]
– Initiated using
INITiate:CONTinuous ON
– Measured using
MEASure?
– Read using
READ?
– If
OFF
is specified, the FCO is not cleared by the above commands.
– If
ONCE
is specified, the FCO is cleared only after the first initialization then
starts accumulating any limit failures.
Syntax
0|OFF
1|ON
Space
?
:LIM
:CLE
:AUTO
ONCE
CALC
1
2
3
4