10
SENSe Subsystem
416
Keysight N8262A Programming Guide
SENSe[1]|2:TRACe:Y:SCALe:PDIV <numeric_value>
This command is used to set the Y Scale value (per division) for the selected trace.
The Y Scale value set by this SCPI command is dependant on the current Y-axis
unit, which can be set using
SENS:TRAC:UNIT
command.
Syntax
Parameters
NOTE
This command is only applicable when N192x sensor is present and trigger
acquisition mode is selected.
1
2
Space
?
:Y
SENS
:TRAC
:SCAL
:PDIV
numeric_value
Item
Description/Default
Range of Values
numeric_value
The numeric value for Y-axis scale.
0.001 dB to10 dB
1 nWatt to10 MWatt