3
CALCulate Subsystem
212
Keysight N8262A Programming Guide
CALCulate[1]|2|3|4:MATH[:EXPRession] <string>
This command sets the specified window/measurement to a single channel,
difference or ratio measurement.
The command may result in a change to the measurement mode set by
CALC:FEED <string>
. The following sequence of commands provides an
example:
1
SENS2:DET:FUN=AVERage
2
CALC:MATH “(SENS1)”
3
CALC:FEED1 “POW:PEAK”
4
CALC:MATH “(SENS2)”
The
FEED1
measurement mode, set in step 3, is made invalid by step 4 and
automatically changed to “
POW:AVER
”.
Syntax
2
:EXPR
:MATH
?
Space
string
CALC
1
3
4