Power Meter Remote Operation
1
Keysight N8262A Programming Guide
85
Auto Averaging Mode
To enable and disable auto filter mode, use the following command:
[SENSe[1]]|SENSe2:AVERage:COUNt:AUTO <boolean>
When the auto filter mode is enabled, the power meter automatically sets the
number of readings averaged together to satisfy the filtering requirements for
most power measurements. The number of readings averaged together depends
on the resolution and the power level currently being measured.
lists
the number of readings averaged for each range and resolution when the power
meter is in auto filter mode.
Figure 1-2
Typical averaged readings on 8480 Series sensors
NOTE
applies to 8480 Series sensors only.
10 dB
10 dB
Minimum Sensor Power
Maximum Sensor Power
Pow
er Senso
r
Dyna
mic Ran
ge
10 dB
10 dB
1
2
3
4
8
8
128
128
Resolution Setting
Numb
er of Aver
ages
1
1
16
256
1
1
2
32
1
1
1
16
1
1
1
8
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...