CALCulate Subsystem
206
Keysight N8262A Programming Guide
CALCulate[1]|2|3|4:LIMit:UPPer[:DATA] <numeric_value>
This command enters a value for the upper test limit for the specified window/
measurement used in the
CALCulate[1]|2|3|4:LIMit:FAIL? test
. The units
used are dependent on the current setting of
UNIT:POWer
and
CALCulate:RELative:STATe
as shown in
. When the measured power is
greater than the value specified in
CALCulate[1]|2|3|4:LIMit:UPPer[:DATA],
CALCulate[1]|2|3|4:LIMit:FAIL?
reports a fail. When the measured level is less
than or equal to the limit, a fail is not reported.
Syntax
Table 3-2
Measurement units
Measurement Mode
Measurement Type
CALC:REL:STAT OFF
CALC:REL:STAT ON
Linear
Log
Linear
Log
Single Channel
Avg, Pk
Watt
dBm
%
dB
Pk–Avg
%
dB
%
dB
Ratio
Avg, Pk, Pk–Avg
%
dB
%
dB
Difference
Avg, Pk
Watt
dBm
%
dB
Pk–Avg
%
dB
%
dB
Space
?
numeric_value
DEF
MIN
MAX
MIN
MAX
Space
:LIM
:UPP
:DATA
CALC
1
2
3
4
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...