SENSe Subsystem
10
Keysight N8262A Programming Guide
363
Query
[SENSe[1]]|SENSe2:BANDwidth|BWIDth:VIDeo?
The query returns the current sensor bandwidth setting.
Query Example
Error Messages
– If the command is used when a sensor other than a P-Series power sensor is
connected, error –241, “Hardware missing” occurs
– If the command is used when a P-Series sensor is connected and set to
AVERage
mode rather than
NORMal
mode, the error –221, “Settings Conflict”
occurs.
SENS2:BAND:VID?
This command queries the current sensor
bandwidth setting for Channel B.
NOTE
Selection of video bandwidth to
LOW, MED
or
HIGH
implements digital signal
processing to ensure a flat bandwidth up to the frequency shown, bandwidths
are flat to ±0.1 dB. In the
OFF
state no corrections are applied and the response
has a slow roll-off.