STATus Subsystem
11
Keysight N8262A Programming Guide
427
Device Status Register Sets
The status registers contain information which give device status information. The
contents of the individual registers of these register sets may be accessed by
appending the commands listed in
“Status Register Set Commands”
.
The following command descriptions detail the SCPI register you require to
control but do not detail the register set commands.
The one device status register set is:
STATus:DEVice:
The following bits in these registers are used by the power meter:
The Channel A and B sensor connected bits (bits 1 and 2), when queried with the
STATus:DEVice:CONDition?
query are set to:
– 1, when a power sensor is connected
– 0, when no power sensor is connected
The Channel A and B sensor connected bits (bits 1 and 2), when queried with the
STATus:DEVice:EVENt?
query indicate whether a power sensor has been
connected or disconnected depending on the state of the corresponding bits of
STATus:DEVice:NTRansition
and
STATus:DEVice:PTRansition
. If the
corresponding bit in:
Bit Number
Decimal Weight
Definition
0
-
Not used
1
2
Channel A sensor connected
2
4
Channel B sensor connected
3
8
Channel A sensor error
4
16
Channel B sensor error
7-15
-
Not used
14
16384
Soft front panel button click
15
-
Bit 15 always 0
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...