Sensor
Different measurement rates are achievable depending on the sensor type being
used, as shown in
Trigger Mode
The power meter has a very flexible triggering system. For simplicity, it can be
described as having three modes:
– Free Run: When a channel is in Free Run, it continuously takes measurements
on this channel. A channel is in free run when
INITiate:CONTinuous
is set to
ON
and
TRIGger:SOURce
is set to
IMMediate
.
– Triggered Free Run: When a channel is in Triggered Free Run Continuous
Trigger, it takes a new measurement each time a trigger event is detected. A
channel is in Triggered Free Run Continuous Trigger when
INITiate:CONTinuous
is set to
ON
and
TRIGger:SOURce
is not set to
IMMediate
.
– Single Shot: When a channel is in Single Shot, it takes a new measurement
when a trigger event is detected and then returns to the idle state. A channel is
in Single Shot when
INITiate:CONTinuous
is set to
OFF
. Note that a
measurement can take several INT/EXT triggers depending on the filter
settings. Refer to
“TRIGger[1]|2:DELay:AUTO <boolean>”
on page 580 for
further information.
Table 1-4
Model of sensor and measurement rates
Sensor
Measurement Rate
NORMal
DOUBle
FAST
8480 Series
20 reading/s
40 reading/s
NA
E-Series E4410 and E9300
50 ms
20 reading/s
25 ms
40 reading/s
Up to 400
P-Series 50
ms
20 reading/s
25 ms
40 reading/s
Up to 1500
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...