Power Meter Remote Operation
86
Keysight N8262A Programming Guide
Illustrates part of the power sensor dynamic range hysteresis.
Figure 1-3
Averaging range hysteresis
Filter Length
You specify the filter length using the following command:
[SENSe[1]]|SENSe2:AVERage:COUNt <numeric_value>
The range of values for the filter length is 1 to 1024. Specifying this command
disables automatic filter length selection. Increasing the value of the filter length
reduces measurement noise. However, the time to take the measurement is
increased.
Range
The power meter has no internal ranges which can be set. The only ranges that
can be set are those of the E-Series power sensor. With an E-Series power sensor
the range can be set either automatically or manually. Use autoranging when you
are not sure of the power level you will be measuring.
Minimum Sensor Power
Minimum Sensor Power + 10 dB
Range Hysteresis
10.5 dB
9.5 dB