1
Power Meter Remote Operation
106
Keysight N8262A Programming Guide
Saving and Recalling Power Meter Configurations
To reduce repeated programming, up to ten power meter configurations can be
stored in the power meter’s non-volatile memory. The error list, remote addresses,
sensor calibration table data, zeroing and calibration information are not stored.
How to Save and Recall a Configuration
Power meter configurations are saved and recalled with the following commands:
*SAV <NRf>
*RCL <NRf>
The range of values for
<NRf>
in the above commands is 1 to 10.
Example Program
10 ASSIGN @POWER TO 713
20 !Configure the power meter
30 OUTPUT @POWER;“UNIT:POW W”
40 OUTPUT @POWER;“SENS:CORR:LOSS2 -10”
50 OUTPUT @POWER;“SENS:CORR:LOSS2:STAT ON”
60 !Save the configuration
70 OUTPUT @POWER;“*SAV 5”
80 PRINT “Configuration Saved”
90 !Now reset the power meter
100 OUTPUT @POWER;“*RST”
110 ! Recall the configuration
120 OUTPUT @POWER;”*RCL 5”
130 PRINT “Configuration Recalled”
140 PRINT “Save and Recall complete”
150 END
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...