14
TRIGger Subsystem
578
Keysight N8262A Programming Guide
INITiate[:IMMediate]:SEQuence[1]|2
This command sets the power meter in the wait for trigger state. When a trigger is
received, the measurement is taken and the result placed in the power meter
memory. If
TRIGger:SOURce
is set to
IMMediate
the measurement begins as soon
as
INITiate:IMMediate
is executed.
Use
FETCh?
to transfer a measurement from memory to the output buffer. Refer to
on page 145 for further information.
Syntax
Example
Error Messages
If the power meter is not in the “idle” state or
INITiate:CONTinuous
is
ON
, error
–213, “INIT ignored” occurs.
NOTE
This command performs the same function as
INITiate[1]|2:[IMMediate]
.
INIT:IMM:SEQ1
This command places Channel A in the
wait for trigger state.
INIT
:IMM
:SEQ
1
2