12
SYSTem Subsystem
506
Keysight N8262A Programming Guide
[SENS[1]]|SENS2:SWE[1]|2|3|4:OFF:TIME
Gate 1: 0
Gate 2: 0
Gate 3: 750 ns
Gate 4: 0
Delay between trigger point and time
gated period.
[SENS[1]]|SENS2:SWE[1]|2|3|4:TIME
Gate 1: 1.0 µs
Gate 2: 250 ns
Gate 3: 250 ns
Gate 4: 0
Length of time gated period for time gated
measurements.
Trigger setup
TRIG[:SEQ[1]|2]:SOUR
INT1
Trigger source set up and acquisition
mode continuous triggering
INIT:CONT
ON
TRIG[:SEQ]:LEV:AUTO
ON
Disable automatic setting of the trigger
level
TRIG[:SEQ]:LEV
AUTO
Automatic Power level
TRIG[:SEQ]:SLOP
POS
Trigger event recognized on the rising
edge of a signal
TRIG[:SEQ]:DEL
0 s
Delay between recognition of trigger
event and start of a measurement
TRIG[:SEQ]:HOLD
MIN
Trigger holdoff
Range
[a]
[SENS[1]]|SENS2:POW:AC:RANG:AUTO
OFF
Auto range off
[SENS[1]]|SENS2:POW:AC:RANG
UPPER
Range set to upper
Step detection
[SENS[1]]|SENS2:AVER:SDET
0
Step detection disabled
Trace setup
[SENS[1]]|SENS2:TRAC:OFFS:TIME
<numeric_value>
–250 ns
Delay between delayed trigger point and
the start of the trace
[SENS[1]]|SENS2:TRAC:TIME
<numeric_value>
1.5 µs
Length of the trace
[a] The Range setting in
is only applicable for E-Series power sensor.
Table 12-23
RADAR: Power meter presets (continued)
Command
Setting
Comments
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...