CALCulate Subsystem
216
Keysight N8262A Programming Guide
CALCulate[1]|2|3|4:PHOLd:CLEar
This command clears the peak hold value for a specified CALC block so that a new
peak hold value can be set.
Syntax
Example
Error Messages
– If no power sensor is connected, error –241 “Hardware missing” occurs.
– If a sensor, other than an P-Series power sensor, is connected, error –241
“Hardware missing” occurs.
– If
SENS:DET:FUNC
is set to
AVER
or
TRIG:SOUR
is set to
INT1
,
INT2
or
EXT
, error
–221 “Settings conflict” occurs.
NOTE
Clearing the peak hold value for a specified CALC block may affect the peak hold
value of other CALC blocks, depending on the CALC channel set up (set by
CALC:MATH:EXPR
).
CALC2:PHOLd:CLEar
This command clears the peak hold value for
CALC2.
:PHOL
:CLE
CALC
1
2
3
4