7
MEMory Subsystem
310
Keysight N8262A Programming Guide
Syntax
Parameters
Example
Query
MEMory:TABLe:GAIN[:MAGNitude]?
The query returns a list of calibration factor/offset points for the currently selected
table.
MEM
:TABL
:GAIN
:MAGN
numeric_value
Space
,
?
Item
Description/Default
Range of Values
numeric_value
A numeric value for the calibration/ offset factors. The units
are PCT.
1.0 to 150.0
MEM:TABL:SEL "Sensor_1"
MEM:TABL:GAIN 97,99.5,97.4
This command enters a reference calibration
factor of 97 % and calibration factors of 99.5
% and 97.4 % into the sensor calibration
table.