SENSe Subsystem
10
Keysight N8262A Programming Guide
395
[SENSe[1]]|SENSe2:SWEep[1]|2|3|4 Commands
These commands set offset time and time gate length as illustrated in the
following diagram:
Offset time and time gate length values can be set for up to four measurement
gates per channel. Measurement gate number is defined by the numeric value
following the
SWEep
component of the command.
The following commands are detailed in this section:
[SENSe[1]]|SENSe2:SWEep[1]|2|3|4:OFFSet:TIME <numeric_value>
[SENSe[1]]|SENSe2:SWEep[1]|2|3|4:TIME <numeric_value>
Incoming signal
from sensor
Trigger
Point
Defined using:
TRIG:LEVel
TRIG:SLOPe
TRIG:HYSTeresis
Pre-
Trigger
With no
trace:
internally
calculated
time
Post-Trigger
With no trace: internally calculated time
Trigger
Delay
TRIG:DEL
Delayed
Trigger
Point
Offset time:
SENSe:SWEep:OFFSet:TIME
Time gate length:
SENSe:SWEep:TIME
Data Collection Time
NOTE
These commands can only be used with P-Series sensors.
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...