TRIGger Subsystem
14
Keysight N8262A Programming Guide
575
INITiate:CONTinuous:SEQuence[1]|2 <boolean>
This command sets the power meter for either a single trigger cycle or continuous
trigger cycles. A trigger cycle means that the power meter exits the wait for
trigger state and starts a measurement. When entering local mode,
INITiate:CONTinuous
is set to
ON
.
If
INITiate:CONTinuous:SEQuence[1|2] <boolean>
is set to:
–
OFF
, the trigger system remains in the idle state until it is set to
ON
, or
INITiate:IMMediate
is received. Once this trigger cycle is complete the
trigger system returns to the idle state.
–
ON
, the trigger system is initiated and exits the idle state. On completion of
each trigger cycle, the trigger system immediately commences another trigger
cycle without entering the idle state.
Syntax
Example
NOTE
This command performs the same functions as
INITiate[1]|2:CONTinuous
<boolean>
.
INIT:CONT:SEQ2 ON
This command places Channel B in a wait for
trigger state.
INIT
:CONT
0|OFF
1|ON
Space
?
:SEQ
1
2
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...