1
Power Meter Remote Operation
76
Keysight N8262A Programming Guide
Example
The following example sets the expected power value to –50 dBm and the
resolution setting to 3 using the lower level commands. The measurement is a
single Channel A measurement carried out on the lower window.
ABOR1
Aborts Channel A
CALC2:MATH:EXPR "(SENS1)"
Displays Channel A on lower window
SENS1:POW:AC:RANG -50
Sets lower range (E-Series sensors only)
DISP:WIND2:RES 3
Sets the lower window’s resolution to setting 3
INIT1
Causes Channel A to make a measurement
FETC2?
Retrieves the lower window’s measurement