7
MEMory Subsystem
288
Keysight N8262A Programming Guide
MEMory:CATalog:TABLe?
This command is used to list the stored sensor calibration (8480 Series sensors
only) and frequency dependent offset tables.
The power meter returns the data in the form of two numeric parameters and as
many strings as there are stored tables.
<numeric_value>,<numeric_value>{,<string>}
– The first numeric parameter indicates the amount of memory, in bytes, used
for the storage of tables.
– The second parameter indicates the memory, in bytes, available for the
storage of tables.
– Each string parameter returned indicates the name, type and size of a stored
table:
–
<string>,<type>,<size>
–
<string>
indicates the name of the table.
–
<type>
indicates
TABL
for a table.
–
<size>
indicates the size of the table in bytes.
For example, a sample of a response may look like:
1178,10040,"DEFAULT,TABL,14","8481A,TABL,116",
"8482A,TABL,74","8483A,TABL,62"...........
The power meter is shipped with a set of predefined sensor calibration tables. The
data in these sensor calibration tables is based on statistical averages for a range
of Keysight Technologies power sensors. These tables can be edited. The
predefined data is listed in your
User’s Guide
. These power sensors and table
numbers are listed in
.
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...