[SENSe[1]]|SENSe2:POWer:AC:RANGe <numeric_value>
This command is only valid when used with an E-Series Power Sensor. Its purpose
is to select one of two power ranges.
– If 0 is selected, the power sensor’s lower range is selected
– If 1 is selected, the power sensor’s upper range is selected
Setting a range with this command automatically switches
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO
to
OFF
.
Syntax
Example
Reset Condition
On reset, the upper range is selected.
Query
[SENSe[1]]|SENSe2:POWer:AC:RANGe?
The query enters a 1 or 0 into the output buffer indicating the status of the power
sensor’s range.
– 1 is returned when the upper range is selected
– 0 is returned when the lower range is selected
POW:AC:RANG 0
This command sets the power sensor to it’s
lower range.
SENS
1
:AC
POW
SENS2
:
:RANG
Space
?
numeric_value