3
CALCulate Subsystem
190
Keysight N8262A Programming Guide
Parameters
Example
Reset Condition
On reset, data_handle is set to :
POW:AVER
.
Item
Description
Range of Values
string
The input measurement type to be fed to the specific input on the CALC block:
–
PEAK
: peak power
–
PTAV
: peak to average
–
AVER
: average
Values may be followed by
ON SWEEP[1]|2|3|4
where the numeric specifies
the gate to be used for the feed. For example: “
POW:PEAK ON SWEEP2
”.
If ON
SWEEP[1]|2|3|4
is not supplied, the gate used is left unchanged.
A feed of “” (empty string) disables the CALC block and switches off that
display line.
“POW:PEAK”
“POW:PTAV”
“POW:AVER”
CALC3:FEED2 “POW:AVER ON
SWEEP2”
This command selects the input for FEED2 of
CALC block CALC3 to be average power,
using gate 2. The channel from which the
feed is taken is determined by
CALC:MATH:EXPR
..