Power Meter Remote Operation
1
Keysight N8262A Programming Guide
93
Using STATus
You can use the
STATus
subsystem to generate an
SRQ
to interrupt your program
when a limit failure occurs. This is a more efficient method than using
SENSe
or
CALCulate
, since you do not need to check the limit failures after every power
measurement.
Refer to
on page 419 for
further information.
NOTE
If
TRIGger:DELay:AUTO
is set to
ON
, then the number of failures returned by
SENSe:LIMit:FCOunt?
or
CALCulate[1|2]:LIMit:FCOunt?
is affected by the
current filter settings.