Power Meter Remote Operation
100
Keysight N8262A Programming Guide
Status Reporting
Status reporting is used to monitor the power meter to determine when events
have occurred. Status reporting is accomplished by configuring and reading
status registers.
The power meter has the following main registers:
– Status Register
– Standard Event Register
– Operation Status Register
– Questionable Status Register
– Device Status Register
There are other registers that exist “behind” the main registers, and are described
later in this chapter.
Status and Standard Event registers are read using the IEEE-488.2 common
commands.
Operation and Questionable Status registers are read using the SCPI
STATus
command subsystem.
The General Status Register Model
The generalized status register model shown in
is the building block of
the SCPI status system. This model consists of a condition register, a transition
filter, an event register and an enable register. A set of these registers is called a
status group.
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...