SERVice Subsystem
16
Keysight N8262A Programming Guide
615
SERVice:BIST:CALibrator <boolean>
This command enables/disables the calibrator self-test during power-up. It can be
used to disable the self-test if it incorrectly indicates a failure. If a load, for example,
a sensor, is connected to the calibrator port this could cause the self-test to fail.
Also, if it fails the self-test, a Pop-up is displayed for 5 seconds, stating -
If Ref
Calibrator test fails disconnect any load attached to it and re-try test
.
Syntax
Example
Query
SERVice:BIST:CALibrator?
The query enters a 1 or 0 into the output buffer indicating the status of the
self-test.
– 1 is returned when the self-test is enabled
– 0 is returned when the self-test is disabled
Query Example
SERV:BIST:CAL OFF
This command disables the calibrator
self-test during power-up.
:BIST
SERV
0|OFF
1|ON
Space
?
:CAL
ONCE
SERV:BIST:CAL?
This command queries whether the self-test
is enabled or disabled.
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...