12
SYSTem Subsystem
512
Keysight N8262A Programming Guide
Table 12-29
802.11b/g: Power meter presets
Command
Setting
Comments
Frequency
[SENS[1]]|SENS2:FREQ[:CW|:FIX]
+2.400 GHz
Frequency setting
Sensor measurement mode
[SENS[1]]|SENS2:DET:FUNC
NORM
Measurement mode
Sensor video bandwidth setup
[SENS[1]]|SENS2:BAND|BWID:VID
N1921/2A: HIGH
Sensor video bandwidth
Gate Setup
[SENS[1]]|SENS2:SWE[1]|2|3|4:OFF:TIME
Gate 1: 0
Gates 2 - 4: 0
Delay between trigger point and time
gated period.
[SENS[1]]|SENS2:SWE[1]|2|3|4:TIME
Gate 1: 100 µs
Gates 2 - 4: 0
Length of time gated period for time gated
measurements.
Trigger setup
TRIG[:SEQ[1]|2]:SOUR
INT1
Trigger source set up and acquisition
mode continuous triggering
INIT:CONT
ON
TRIG[:SEQ]:LEV:AUTO
ON
Enable automatic setting of the trigger
level
TRIG[:SEQ]:LEV
AUTO
Power level
TRIG[:SEQ]:SLOP
POS
Trigger event recognized on the rising
edge of a signal
TRIG[:SEQ]:DEL
0 s
Delay between recognition of trigger event
and start of a measurement
TRIG[:SEQ]:HOLD
MIN
Trigger holdoff
Rang
e
[a]
[SENS[1]]|SENS2:POW:AC:RANG:AUTO
OFF
Auto range off
[SENS[1]]|SENS2:POW:AC:RANG
UPPER
Range set to upper
Step detection
[SENSe[1]]|SENS2:AVER:SDET
0
Step detection disabled
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...