MEMory Subsystem
7
Keysight N8262A Programming Guide
313
MEMory:TABLe:MOVE <character_data>,<character_data>
This command is used to rename a sensor calibration table (8480 Series Sensors
only) or a frequency dependent offset table.
Syntax
Parameters
Example
Error Messages
– If either table name is invalid, error –224, “Illegal parameter value” occurs.
– If the first parameter does not match an existing table name, error –256, “File
name not found” occurs.
– If the second parameter matches an existing table name or save/recall
register, error –257, “File name error” occurs.
MEM
:TABL
:MOVE
,
Space
character_data
character_data
Item
Description/Default
Range of Values
character_data
(1st parameter)
Contains the existing table name.
existing table name
character_data
(2nd parameter)
Details the new table name. A maximum of 12 characters
can be used.
A to Z (uppercase)
a to z (lowercase)
0 - 9
_ (underscore)
MEM:TABL:MOVE "tab1","tab1a"
This command renames a table named tab1
to tab1a.
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...