13
TRACe Subsystem
550
Keysight N8262A Programming Guide
TRACe[1]|2:MEASurement:PULSe[1]|...|10:PERiod?
This command returns the pulse period. This is the time difference between two
consecutive transition occurrences of the same polarity. The period is equal to the
sum of the pulse separation and the pulse duration.
Algorithm
If the first transition in the trace is positive,
then
PULSe:PERiod
=
time the second positive transition occurrence
-
time the first
positive transition occurs
else
PULSe:PERiod
=
time the second negative transition occurs
-
time the first
negative transition occurs
.
Syntax
Example
TRAC
:MEAS
1
2
:PER
?
1
:PULS
2|...|10
TRAC:MEAS:PULS:PER?
This command returns the period of the
pulse found on trace 1.