CALibration Subsystem
4
Keysight N8262A Programming Guide
229
Syntax
Query Example
Error Messages
– If the calibration was not carried out successfully the error –231, “Data
Questionable; CAL ERROR” occurs. The error message will specify which
channel failed calibration.
– If zeroing was not carried out successfully the error –231, “Data Questionable;
ZERO ERROR” occurs.
The error message will specify which channel failed
calibration.
– If there is no sensor connected, the error –241, “Hardware Missing” occurs.
CAL1:ALL?
This command causes the power meter to
perform a calibration sequence on Channel A
and return a result.
:ALL
?
CAL
1
2