SENSe Subsystem
10
Keysight N8262A Programming Guide
347
[SENSe[1]]|SENSe2:AVERage:COUNt <numeric_value>
This command is used to enter a value for the filter length. If
[SENSe[1]]|SENSe2:AVERage:COUNt:AUTO
is set to
ON
then entering a value for
the filter length automatically sets it to
OFF
. Increasing the value of filter length
increases measurement accuracy but also increases the time taken to make a
power measurement.
Entering a value using this command automatically turns the
[SENSe[1]]|SENSe2:AVERage:STATe
command to
ON
.
NOTE
For most applications, automatic filter length selection
(
[SENSe[1]]|SENSe2:AVERage:COUNt:AUTO ON
) is the best mode of operation.
However, manual filter length selection (
[SENSe[1]]|SENSe2:AVERage:COUNt
<numeric_value>
) is useful in applications requiring either high resolution or fast
settling times, where signal variations rather than measurement noise need
filtering, or when approximate results are needed quickly.
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...