7
MEMory Subsystem
292
Keysight N8262A Programming Guide
MEMory:CLEar[:NAME] <character_data>
This command clears the contents of a specified sensor calibration table (8480
Series sensors only), frequency dependent offset table, or save/recall register.
Although the table remains, a
MEMory:TABLe:FREQuency|GAIN:POINts?
query
returns a 0 as there are no contents in the table.
For sensor calibration tables and frequency dependent offset tables, this
command is an alternative form of the
MEMory:CLEar:TABLE
command, the only
difference being the method in which the table is selected.
NOTE
The contents cleared using this command are non-recoverable.