5
DISPlay Subsystem
256
Keysight N8262A Programming Guide
Parameters
Example
Reset Condition
On reset, the defaults for the upper and lower windows are
DIGital
.
Item
Description/Default
Range of Values
character_data
Sets the window format:
–
DIGital
: sets the window display to digital. This setting is
the same as SNUMeric.
–
ANALog
: sets the window display to analog using the
currently SELected measurement.
–
SNUMeric
: sets the window display to single numeric. The
currently SELected measurement is displayed. This setting is
the same as DIGital.
–
DNUMeric
: sets the window display to dual numeric.
–
TRACe
: trace display using the currently SELected
measurement. Used to determine the channel from which
the trace is taken.
–
CTABle
: sets the window display to CCDF table.
DIGital
ANALog
SNUMeric
DNUMeric
TRACe
CTABle
DISP:WIND2:FORM DIG
This command sets the lower window to a
digital display.
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...