Example 6 - Making a Ratio Measurement
The following command is performed on the N8262A. It queries the upper window
to make a ratio measurement of Channel A/B. The expected power and resolution
parameters are defaulted, leaving them at their current settings.
MEAS1:POW:AC:RAT? DEF,DEF,(@1),(@2)
specifies window
specifies the relationship of the
channels in the ratio
Channel A / B
NOTE
Keysight N8262A
The operation of the
MEASure?
command when the source list parameter is
defaulted depends on the current setup of the window concerned (for example,
A, B, A/B, A-B etc.) and on the particular command used (for example,
MEAS[:POW][:AC]?
and
MEAS:POW:AC:RAT?
).
This means that when the source list parameter is defaulted, there are a number
of possibilities.