9
PSTatistic Subsystem
338
Keysight N8262A Programming Guide
Syntax
Example
PST
1
:CCDF
:TAB
2
?
PST:CCDF:TAB?
This command returns the data in CCDF table:
average input power, probabilty at the average
input power, power level at various predefined
probability steps (10 %, 1 %, 0.1 %, 0.01 %,
0.001 % and 0.0001 %), peak to average power
ratio and sample count for Channel A.
PST2:CCDF:TAB?
This command returns the data in CCDF table:
average input power, probabilty at the average
input power, power level at various predefined
probability steps (10 %, 1 %, 0.1 %, 0.01 %,
0.001 % and 0.0001 %), peak to average power
ratio and sample count for Channel B.
NOTE
The sample count will always be returned as 100 million samples (100,000,000).