1
Power Meter Remote Operation
70
Keysight N8262A Programming Guide
The following example uses the expected value parameter to specify a value of
–50 dBm. This selects the power meter’s lower range (refer to
for details of the range breaks). The resolution parameter is defaulted, leaving it at
its current setting. The source list parameter specifies a Channel B measurement.
The measurement is carried out on the upper window.
Using READ?
Some fine tuning of measurements can be performed using the
CONFigure
and
READ?
commands. For example, in the above program segment some fine tuning
can be performed by setting the filter length to 1024 and the trigger delay off.
1
ABOR2
2
CONF1 -50,DEF,(@2)
3
SENS2:AVER:COUN 1024
4
TRIG2:DEL:AUTO OFF
5
READ1?
Using INITiate and FETCh?
ABOR2
Aborts Channel B
CONF1 -50,DEF,(@2)
Configures the upper window to make a
Channel B measurement using an expected
power of –50 dBm and the current resolution
setting
READ1?
Takes the upper window’s measurement
ABOR2
Aborts Channel B
CONF1 -50,DEF,(@2)
Configures the upper window to make a Channel
B measurement using an expected power of
–50 dBm and the current resolution setting
INIT2
Causes Channel B to make a measurement
FETC1? -50,DEF,(@2)
Retrieves the upper window’s measurement
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...