MEASurement Commands
148
Keysight N8262A Programming Guide
FETCh[1]|2|3|4[:SCALar][:POWer:AC]:RELative?
[<expected_value>[,<resolution>[,<source list>]]]
This command sets the specified window’s measurement function to single
channel with relative mode on, recalculates the measurement and places the
results on the bus. The result is a ratio based measurement and is expressed in the
units defined by
UNIT[1]|2|3|4:POWer:RATio
. The relative value used is that set
by the
CALCulate:RELative:MAGNitude:AUTO
command.
Syntax
Parameters
Refer to
on page 126 for additional details on the
parameters in this command.
:POW
:AC
?
:SCAL
:REL
Space
expected_value
DEF
resolution
,
DEF
source list
,
FETC
1
2
3
4
Item
Description/Default
Range of Values
expected_value
(for the expected power
level)
The expected power level parameter can be set to
DEF
or a
numeric value. If a value is entered it should correspond to that set
by
CONFigure
otherwise an error occurs. The units of
measurement are dBm and W. The default units are defined by
UNIT:POWer
.
Sensor dependent
DEF
[a]
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...