TRIGger Subsystem
14
Keysight N8262A Programming Guide
603
Parameters
Example
Reset Condition
On reset, the trigger source is set to
IMMediate
.
Item
Description/Default
Range of Values
source
– Available trigger sources:
–
BUS
: the trigger source is the group execute trigger
<GET>
bus command, a
*TRG
common command or the
TRIGGER:IMMediate
SCPI command.
–
EXTernal
: the trigger source is the trigger input in the front panel.
–
HOLD
: triggering is suspended. The only way to trigger the power meter is to use
TRIGger:IMMediate
.
–
IMMediate
: the trigger system is always true. If
INITiate:CONTinuous
is
ON
the power meter is continually triggering free (free run mode). If an
INITiate:IMMediate
command is sent a measurement is triggered then the
power meter returns to the idle state.
–
INTernal
: either
INT1
(Channel A) or
INT2
(Channel B).
BUS
EXTernal
HOLD
IMMediate
INTernal[[1]|2]
NOTE
TThe trigger source is set to
IMMediate
on instrument power-up and when
entering local mode.
The
MEASure
and
CONFigure
commands automatically set the trigger source to
IMMediate
.
The
READ?
or
MEASure
commands should not be used if the trigger source is set
to
BUS
or
HOLD
.
TRIG:SOUR IMM
This command configures Channel A for
immediate triggering.
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...