10
SENSe Subsystem
358
Keysight N8262A Programming Guide
[SENSe[1]]|SENSe2:AVERage2:COUNt <numeric_value>
This command is used to enter the video filter length for the P-Series sensor.
Video filtering is applied to the traces. Successive traces are combined to reduce
noise without affecting the dynamic characteristic of the signal.
Syntax
Parameters
Example
Reset Condition
On reset, the filter length is set to 4.
:COUN
Space
?
numeric_value
DEF
SENS
1
AVER2
SENS2
:
Item
Description/Default
Range of Values
numeric_value
A numeric value defining the filter length.
–
DEF
: the default value is 4.
1 to 256
[a]
DEF
[a] This is only implemented in powers of 2 (2
n
).
AVER2:COUN 16
This command enters a video filter length of
16 for Channel A.
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...