Query Example
Error Messages
– If
[SENSe[1]]|SENSe2:CORRection:DCYCle:STATus
is set to
ON
while
[SENSe[1]]|SENSe2:SPEed
is set to 200, the error –221, “Settings Conflict”
occurs.
– If this command is used when an P-Series or E-Series Power Sensor is
connected, the error –310, “System error;Dty Cyc may impair accuracy with
ECP sensor” occurs. If you are using a dual channel power meter the error
message specifies the channel.
CORR:GAIN3:STAT?
This command queries whether the pulse
measurement feature is on or off.