2
MEASurement Commands
134
Keysight N8262A Programming Guide
Example
resolution
A numeric value for the resolution. If unspecified the current
resolution setting is used.
1 to 4
[b]
1.0, 0.1, 0.01, 0.001
source list
The channel which the command is implemented on.
If unspecified the current window setup is used. However, if
the window shows a ratio or difference measurement, the
upper window defaults to Channel A and the lower window
to Channel B.
(@1)
(@2)
[a] The mnemonic
DEF
means DEFault. This is not equivalent to the DEFault parameter used in the command sub-systems. The parameters must
be entered in the specified order. If parameters are omitted, they default from the right. The parameter DEFault is used as a place holder.
Specifying
DEF
leaves the parameter value unchanged.
[b] When the measurement result is linear this parameter represents the number of significant digits. When the measurement result is
logarithmic 1 to 4 represents of 1, 0.1, 0.01 and 0.001 respectively.
CONF1:POW:AC DEF,2,(@1)
This command configures the upper window/
upper measurement to measure the power of
Channel A, using the current sensor range
and a resolution setting of 2.
Item
Description/Default
Range of Values
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...