Power Meter Remote Operation
1
Keysight N8262A Programming Guide
61
Making Measurement
The
MEASure?
and
CONFigure
commands provide a straight-forward method to
program the power meter for measurements. You can select the measurement’s
expected power level, resolution and with the N8262A the measurement type
(that is single channel, difference or ratio measurements) all in one command. The
power meter automatically presets other measurement parameters to default
values as shown in
An alternative method to program the power meter is to use the lower level
commands. The advantage of using the lower level commands over the
CONFigure
command is that they give you more precise control of the power
meter. As shown in
CONFigure
command presets various states in
the power meter. It may be likely that you do not want to preset these states.
“Using the Lower Level Commands”
on page 75 for further information.
Table 1-1
MEASure? and CONFigure preset states
Command
MEASure? and CONFigure Setting
Trigger source
(
TRIGger:SOURce
)
Immediate
Filter
(
SENSe:AVERage:COUNt:AUTO
)
On
Filter
state(
SENSe:AVERage:STATe
)
On
Trigger cycle
(
INITiate:CONTinuous
)
Off
Trigger Delay
(
TRIGger:DELay:AUTO
)
On
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...