SENSe Subsystem
10
Keysight N8262A Programming Guide
413
Parameters
Example
Reset Condition
On reset, the value is set to
DEF
.
Query
SENSe[1]|2:TRACe:LIMit:LOWer [MIN|MAX]
The query returns the current setting of the trace upper scale limit or the value
associated with
MIN
or
MAX
. The format of the response is
<NR3>
.
Query Example
Item
Description/Default
Range of Values
numeric_value
A numeric value for the trace lower scale limit.
–
DEF
: the default is 20 dBm
–
MIN
: –150 dBm
–
MAX
: 230 dBm
–150 to 230 dBm
DEF
MIN
MAX
SENS:TRAC:LIM:UPP 100
This command sets the trace upper scale
limit to 100 dBm for Channel A.
SENS:TRAC:LIM:UPP?
This command queries the trace upper scale
limit of Channel A.