SENSe Subsystem
350
Keysight N8262A Programming Guide
[SENSe[1]]|SENSe2:AVERage:COUNt:AUTO <boolean>
This command enables and disables automatic averaging.
ONCE
has no affect on
the power meter.
When the auto filter mode is enabled, the power meter automatically sets the
number of readings averaged together to satisfy the averaging requirements for
most power measurements. The number of readings averaged together depends
on the resolution and the power level in which the power meter is currently
operating.
is an example of the averaged number of readings for each
range and resolution when the power meter is in auto measurement average
mode and using a P-Series
power sensor.
Setting this command to
ON
automatically sets the
[SENSe[1]]|SENSe2:AVERage:STATe
command to
ON
.
Figure 10-1
Example of averaged readings
If
[SENSe[1]]|SENSe2:AVERage:COUNt:AUTO
is set to
OFF
, the filter length is set
by the
[SENSe[1]]|SENSe2:AVERage:COUNt
command. Using the
[SENSe[1]]|SENSe2:AVERage:COUNt
command disables automatic averaging.
Auto averaging is enabled by the
MEASure:POWer:AC?
and
CONFigure:POWer:AC?
commands.
10 dB
10 dB
Maximum Sensor Power
Minimum Sensor Power
Pow
er Sens
or
Dynamic Range
10 dB
10 dB
1
2
3
4
1
1
1
8
Resolution Setting
Number
of
A
ver
ag
es
1
1
1
16
1
1
2
32
1
1
16
256
1
8
128
128
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...