1
Power Meter Remote Operation
88
Keysight N8262A Programming Guide
Setting Offsets
Channel Offsets
The power meter can be configured to compensate for signal loss or gain in your
test setup (for example, to compensate for the loss of a 10 dB attenuator). You use
the
SENSe
command subsystem to configure the power meter. Gain and loss
correction are a coupled system. This means that a gain set by
[SENSe[1]]|SENSe2:CORRection:GAIN2
is represented in the
[SENSe[1]]|SENSe2:CORRection:LOSS2?
command. If you enter an offset value
the state is automatically enabled. However it can be enabled and disabled using
either the
[SENSe[1]]|SENSe2:CORRection:GAIN2:STATe
or
[SENSe[1]]|SENSe2:CORRection:LOSS2:STATe
commands.
LOSS2
is coupled to
GAIN2
by the equation
when the default unit is
linear, and
when the default is logarithmic.
Display Offsets
Display offset values can be entered using the
CALCulate[1|2]:GAIN[:MAGNitude]
command.
CALCulate[1|2]:GAIN:STATe
must be set to
ON
to enable the offset value. If you enter an offset value the state is
automatically enabled. This offset is applied after any math calculations (refer to
Example
The following example program, in HP Basic, details how to use the channel and
display offsets on an N8262A making a Channel A/B ratio measurement.
The final result is:
Loss
1
Gain
---------
=
Gain
Loss
–
=
NOTE
You can only use
LOSS2
and
GAIN2
for external losses and gains.
LOSS1
and
GAIN1
are specifically for calibration factors.
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...