Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2461-901-01 A/November 2015
8-191
Example
smu.source.func = smu.FUNC_DC_VOLTAGE
smu.source.configlist.create("CustomVoltSweep")
smu.source.pulse.level = 1.17
smu.source.configlist.store("CustomVoltSweep")
smu.source.pulse.level = 1.27
smu.source.configlist.store("CustomVoltSweep")
smu.source.pulse.level = 1.33
smu.source.configlist.store("CustomVoltSweep")
smu.source.pulse.level = 1.39
smu.source.configlist.store("CustomVoltSweep")
smu.source.pulsesweeplist("CustomVoltSweep", 0.03, smu.OFF, defbuffer1, 1, 3, 0,
.005)
Set the source function to voltage and create a source list called
CustomVoltSweep
.
Set the pulse level to 1.17 V and store that pulse level at the next index (1) in
CustomVoltSweep
.
Set the pulse level to 1.27 V and store that pulse level at the next index (2) in
CustomVoltSweep
.
Set the pulse level to 1.33 V and store that pulse level at the next index (3) in
CustomVoltSweep
.
Set the pulse level to 1.39 V and store that pulse level at the next index (4) in
CustomVoltSweep
.
Creates a pulse sweep using the
CustomVoltSweep
source list that steps through four pulse levels three
times, for a total of 12 pulses. Each pulse has a width of 30 ms and an off time of 5 ms. Readings are not
taken during this pulse sweep. Start the sweep at index 1. Then query the pulse list.
Output:
1.17,1.00,1.00,1.00